7 CH1 ON/OFF & Amplitude Control Physioworx E-Pulse 300 with Probe CH2 ON/OFF & Amplitude Control CH1 Output Receptacle CH2 Output Receptacle LCD Panel Key Control (Rate, Width, Mode & Time) Probe Channel: Dual, isolated between channels. Pulse intensity:Adjustable0-80mApeakinto500Ωloadeachchannel, constant current (Contraction mode). Normal mode of Obstetric: 80%90% amplitude of setting. Pulse Rate: Contraction 1-160Hz (adjustable) , 1Hz/step (1-20Hz), 5Hz/ step (20-160Hz). Normal: 2 Hz fixed. Pulse Width: 50μs - 260μs (adjustable), 10μs/step. Patient Compliance meter: Shows the treatment times in hours. Patient Lock System: Prevent the user from changing any fixed parameters set by the physician or licensed practitioner. Software Ramp Up: By change mode, the output will reset to zero; then ramp up to its original setting intensity. This will protect users from a sudden surge. Timer: 15, 30, 60 minute and continuous mode. LCD: Shows modes, pulse rate, pulse width, timer, CH1/CH2. Weight: 140 grams (battery included). (L)108mm x (W)61.5mm x (H)25mm PX0400 Hand control probe input Receptable Physioworx Self Adhesive Electrodes High quality self adhesive electrodes, replacement pads for use with TENS therapy units. Each pack contains 4 electrodes (Round, Square or Rectangle) PX1964 PX1966 PX1968 2.5cm 4cm 5cm PX1956 PX1958 4 x 4cm 5 x 5cm PX1960 PX1962 4 x 8cm 10 x 5cm Replacement Standard TENS Machine Leadwire R&B These TENS lead wires fit most standard machines on the market. Sold Individually PX1970 Standard